Customization: | Available |
---|---|
Capacity: | One Person |
After-sales Service Provided: | Engineers available to service machinery overseas |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Benchtop biosafety cabinet
Secondary biosafety cabinet, air flow mode: 30% discharge, 70% internal circulation.
It conforms to nsf49 and en12469 standards, and the product executive standard is the pharmaceutical industry standard of the people's Republic of China, YY 0569-2011.
The outer box body is made of high-quality cold-rolled steel plate and electrostatic spray plastic; the three sides of the operation area are made of stainless steel integrated structure, and the inner parts that can be cleaned are made of 10 mm large fillet for easy cleaning.
Two ULPA ultra-high efficiency filters are equipped as standard, which can achieve 99.999% closure efficiency for 0.12 μ m particle system. The filter membrane is made of non barrier borosilicate glass fiber.
The control software system is independently developed and has obtained the national software copyright certificate. The LCD screen can display the system parameters such as the working area temperature, air flow rate, service life of the filter membrane, and can adjust the parameter settings. The blue screen digital indicates the parameter values, and the password modification program settings to prevent misoperation.
Unique negative pressure leak proof design, the work area surrounded by negative pressure, truly achieve zero leakage, to ensure the cleanliness of the operation area.
The fan is equipped with voltage fluctuation compensation function and resistance induction compensation function, effectively extending the service life of ultra-high efficiency filter.
A high-sensitivity and high-precision breeze speed sensor is equipped at the working area and the external air outlet to detect the wind speed in real time.
Temperature sensor: it can detect and display the temperature in real time.
Equipped with filter pre failure, damage and blocking alarm technology, equipped with fan overload alarm prompt.
Equipped with height limit alarm device, when the mobile glass door exceeds the specified height limit (200mm), the user will be alerted by sound and light.
It is equipped with a working window opening limit alarm system. When the opening height of the front window of the safety cabinet exceeds the set height, when the flow velocity of the down airflow and the flow velocity of the inflow airflow of the safety cabinet fluctuates by more than 20% of its nominal value, the interlocking system will start the alarm prompt.
With UV appointment timing function.
The front window integrated sliding sliding door can be positioned arbitrarily, and can be completely closed for sterilization. The front window glass door shall be made of safety toughened glass with a thickness of no less than 6mm.
Interlock design of front window and germicidal lamp: when the front window is opened, the germicidal lamp is off.
It has the function of parameter memory and call recovery protected by power failure and data loss of dead state.
The front suction inlet adopts the design of unimpeded return air to ensure that the formation of counter current is effectively prevented and the safety is improved during the operation.
10 degree tilt design, more in line with ergonomic power, making the operator more comfortable.
Power supply in the cabinet: waterproof socket design, the whole machine has power-off protection function, making the operation more flexible and convenient.
The manufacturer has relevant qualification and patent
(1)The company has passed ISO9001 quality management system certification.
(2)The manufacturer has obtained the medical device production license.
(3) The product has obtained SFDA medical device registration certificate.
(4) The control software obtained the national software copyright certificate.
(5) The product has dozens of patents.
Before leaving the factory, the equipment shall be tested by a qualified third party to ensure that the equipment is qualified to delivery.
Parameter / model | BSC-1000A2 |
Cleanliness grade | HEPA: ISO5 (Class 100 of 100) / ULPA: iso4 (class 10 of 10) |
Filter level / efficiency | HEPA:≥99.995%,@0.3μm/ULPA:≥99.999%,@0.12μm |
Descent wind speed | 0.33m/s |
Suction air speed | 0.53m/s |
Noise | ≤65dB |
Vibration half peak | ≤5μm |
Power / maximum power consumption | AC220V/50HZ/1KW |
Biosafety personnel protection | CFU of all impact samplers ≤ 10, CFU of slit samplers ≤ 5 |
Product protection | CFU of all sampling dishes ≤ 5 |
Cross contamination protection | CFU of all culture dishes ≤ 2 |
Workspace size | 1000*520*640mm |
Overall dimension of device | 1165*760*1550mm |
Specification of air supply filter | 990*470*70*(1) |
Specification of exhaust air filter | 650*450*90*(1) |
Specification and quantity of fluorescent lamp / UV lamp |
30W*(2)/30W*(1) |
Illuminance | ≥900LX |
Weight | 230kg |
Applicable number | 1 |
Wind direction | Top out |