Product Description
The GGSD71 HTHP dynamic fluid loss tester is a test instrument that simulates the fluid loss of drilling and water drilling fluids in deep wells (high temperature and high pressure). It is suitable for:
1. Static fluid loss and filter cake preparation test: at any temperature from room temperature to 232ºC, the maximum working pressure of the drilling fluid cup is less than 7.1MPa, the filter loss pressure difference is 3.5MPa, and the filter loss area is 22.6cm2.
2. Dynamic fluid loss and filter cake preparation test: at any temperature from room temperature to 150ºC, the maximum working pressure of the drilling fluid cup is less than 7.1MPa, the filter loss pressure difference is 3.5MPa, and the filter loss area is 22.6cm2.
3. There are double screens, which can meet the measurement of cement slurry.
Technical Specifications
1 |
Power supply |
220V±5%,50Hz, |
2 |
Power |
1.2KW |
3 |
Speed range |
~800r/min (stepless adjustable) |
4 |
Static filter loss test working temperature |
≤232ºC |
5 |
Dynamic filter loss test working temperature |
≤120ºC |
6 |
Filter area |
22.6cm2 |
7 |
Safe working pressure of drilling fluid cup |
8MPa |
8 |
Safe working pressure of back pressure device |
4MPa |
9 |
Drilling fluid cup capacity |
500ml |
10 |
Back pressure receiver capacity |
95ml |
11 |
Heating rate |
30 minutes after the power is turned on, the heater temperature rises above 200ºC |
12 |
Heating accuracy |
Heater temperature change after constant temperature <±3ºC, drilling fluid chamber <±2ºC |
13 |
Gas source |
Nitrogen and carbon dioxide gas without impurities such as oil and water. The use of oxygen is strictly prohibited |
Related Product
Low pressure API filter press(ZNS-5A 1-gang)
Low pressure API filter press (SD-6A 6-gang)
Our main drilling fluids & oil well cement instruments
Company Profile