• UV Aging Test Chamber ASTM D7238 ASTM G154 G155

UV Aging Test Chamber ASTM D7238 ASTM G154 G155

After-sales Service: Life Time
Warranty: 12 Months Free
Delivery Time: 10-15 Working Days
Application: UV Radiation Aging Test
Power Supply: AC 220V, 50Hz or Others
Customized: Yes
Customization:
Manufacturer/Factory, Trading Company
Diamond Member Since 2020

Suppliers with verified business licenses

Rating: 5.0/5
Jiangsu, China
In-stock Capacity
The supplier has In-stock capacity
Fast Delivery
The supplier can deliver the goods within 15 days
Low MOQ
The MOQ for the supplier's products is 1
Customization from Samples
The supplier provides sample based customization services
to see all verified strength labels (14)

Basic Info.

Model NO.
CT-UV3-A
Material Made
SUS 304
Controller
Touch Screen LCD Display
Transport Package
Wooden Case
Specification
CT-UV3-A
Trademark
SERVE REAL
Origin
China
HS Code
9031809090
Production Capacity
300sets/Year

Product Description

Standards
ASTM D4329, ISO 4892-3, ISO 11507, SAE J2020, ASTM D7238, ASTM G154, G155

Features
1. Provide comprehensive and detailed real-time monitoring and curve recording for important test parameters such as irradiance, temperature and time (controller with recorder function), and generate reports and curve printing through computer.
2. Irradiation/condensation/water spray test can be conducted separately, or any combination of irradiation/condensation/water spray test can be programmed.
3. The independently developed industrial level 7-inch true color touch screen control system is adopted, with good stability, monitoring, operability and maintainability.
4. The service life of the imported ultraviolet lamp with expensive materials (the lamp is a consumable, and the service life is about 1600 hours) is provided to facilitate the user to confirm the replacement time of the lamp and save the use cost.
5. The automatic water supply and drainage design is adopted, and the high-power pump with high cost is used to ensure the flow rate and uniformity of water spray.
6. The installation and wiring of the electrical part of the equipment shall comply with the international electrical standards.

Specifications

Model CT-UV3-A
Equipment function Lighting, condensation and spraying (three function slicing arrangement)
Inner box size W 114 X H 39 X D 40 cm (W X H X D)
Test piece 75mm * 140mm, 48 pieces
Approximate overall dimensions W 130 X H 148 X D 55 cm (W X H X D)
Temperature range Adjustable at room temperature+10~70 ºC
Humidity range ≥90%RH
Lamps 8pcs UVA-340 (315~400nm) default configuration, 8pcs UVB-313 (280~315nm) (optional)
Lamp brand US Q-Lab or US ATLAS imported with original packaging (100% compensation for one fake)
Center distance of lamp 70mm
Center distance between the test object and the lamp tube: 50±3mm
Irradiation intensity Adjustable or maximum power output within 0.35~0.9W/m2 (adjustable only if the solar eye online control system is included)
Temperature fluctuation ±3ºC
Temperature resolution: 0.1ºC
Blackboard thermometer 45~80ºC; Tolerance: ± 3 ºC
Temperature control mode PID self-tuning temperature control mode
Environmental conditions When the ambient temperature is+28 ºC, the relative humidity is ≤ 85%, and there is no sample in the test chamber


Chamber
Inner box material The inner plate is made of high quality SUS304 stainless steel, with good corrosion resistance, heat resistance, low temperature strength and mechanical properties; Stamping, bending and other hot workability are good, and there is no heat treatment hardening phenomenon (non-magnetic, use temperature - 196 ºC~800 ºC).
Outer box material The box body is made of full numerical control, with advanced technology, smooth lines, and beautiful appearance. The external material can be made of high-quality SUS304 stainless steel or fine powder baking paint.
Equipment sample area Dimension of standard test piece: 75mm * 140mm, 48 pieces, 75mm * 280mm, 24 pieces. (Non standard size can be customized) Schematic diagram of sample suspension at the illumination part of sample rack front


Control system
Controller Self developed 7-inch super large touch intelligent controller: it has fixed value and program options.
Operation mode Manual and programmed control
Input method Touch screen  
Setting method
Program Editing
Constant value operation function: the user can use this function to realize continuous ultraviolet test function. In the customized operation screen, the user can manually control the opening and closing of each function. It can realize temperature adjustment and has the function of automatic temperature calculation.
Program test run function: On the program control screen, the user can automatically run the equipment according to the edited program. When the program is running, the device has the function of keeping skip segments. And it can intuitively observe the total running time of the program, the running time of the program segment, the program number and other basic information. At the same time, the operation status of each system will be displayed on the screen intuitively.
Blackboard temperature monitoring 120 groups of programs can be programmed, and each group of programs can be programmed with a maximum of 100 segments, and cyclic operation can be set;
Communication interface The standard Pt-100 blackboard temperature sensor is used to precisely control the sample surface temperature during the test process.
Ancillary functions (RS232/RS485, maximum communication distance 1.2km [optical fiber up to 30km])
Controller Fault alarm and cause, processing prompt function, power failure protection function, timing function (automatic start and automatic stop), self diagnosis function.



Operation & storage conditions
Installation place The ground is flat and well ventilated, and there is no strong vibration around the equipment
No strong electromagnetic field around the equipment
No flammable, explosive, corrosive substances and dust around the equipment
Proper use and maintenance space shall be reserved around the equipment.
Working conditions Temperature: 5 ºC~ 28 ºC
Relative humidity: ≤ 85%
Air pressure: 86kPa~106kPa
Power supply AC220 V, two wire+protective ground wire
Installed power: 4.5 KW
Allowable voltage fluctuation range: AC (220 ± 10%) V
Allowable frequency fluctuation range: (50 ± 0.5) Hz
The grounding resistance of protective ground wire is less than 4 Ω
Users are required to configure air or power switch with corresponding capacity for the equipment at the installation site, and the switch must be used independently for the equipment
Storage environment When the equipment is not working, the ambient temperature shall be kept within 0 ºC~+45 ºC. When the ambient temperature is lower than 0 ºC, the water in the equipment shall be drained to prevent the water in the pipeline from freezing and damaging the pipeline.
Others Open the door of the test box during the test. Long time exposure to the open lamp will cause UV damage to eyes and skin. Please open the door as little as possible when starting the test.